Half-life measurement of 124Sb

Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1555-60; discussion 1560. doi: 10.1016/j.apradiso.2009.11.050. Epub 2009 Dec 3.

Abstract

The half-life of (124)Sb was determined experimentally by following the decay of a source from a radiopure solution with a Centronic IG12 ionisation chamber. Thousands of measurements were performed over a period of 358 days, i.e. about six half-life periods. However, the data analysis was restricted to the first 221 days, in order to limit the dominant uncertainty component associated with the hypothetical possibility of a systematic error on background subtraction. The resulting value for the (124)Sb half-life, 60.212 (11) days, is found to be in very good agreement with published values, but carries a lower uncertainty. Major uncertainty contributions pertain to possible systematic errors in background correction, long-term changes in source-detector geometry and medium- and long-term instability of the instrument. Additional measurements were performed with a high-purity germanium detector to confirm the above value.

MeSH terms

  • Antimony*
  • Half-Life
  • Radioisotopes*
  • Uncertainty
  • Weights and Measures

Substances

  • Radioisotopes
  • Antimony