National comparison of 131I measurement among nuclear medicine clinics of eight countries

Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1371-7. doi: 10.1016/j.apradiso.2009.11.008. Epub 2009 Nov 20.

Abstract

A generally applicable protocol for organizing comparisons among nuclear medicine clinics created within the IAEA project CRP E2.10.05 was tested in Brazil, Cuba, Czech Republic, India, Iran, Republic of Korea, Romania and Turkey in 2007. Comparisons of measurement of (131)I were organized by local pilot laboratories with different backgrounds and levels of experience in this field. The results and experiences gained were compared and analyzed. A majority of results in each national comparison were within 10% of the reference value.

MeSH terms

  • International Cooperation
  • Iodine Radioisotopes / analysis*
  • Laboratories, Hospital / standards
  • Methods
  • Nuclear Medicine / methods*
  • Nuclear Medicine / standards
  • Observer Variation
  • Quality Assurance, Health Care
  • Radiation Dosage
  • Reference Values
  • Reproducibility of Results

Substances

  • Iodine Radioisotopes