Simple electrical model and initial experiments for intra-body communications

Annu Int Conf IEEE Eng Med Biol Soc. 2009:2009:697-700. doi: 10.1109/IEMBS.2009.5332704.

Abstract

Intra-Body Communication(IBC) is a short range "wireless" communication technique appeared in recent years. This technique relies on the conductive property of human tissue to transmit the electric signal among human body. This is beneficial for devices networking and sensors among human body, and especially suitable for wearable sensors, telemedicine system and home health care system as in general the data rates of physiologic parameters are low. In this article, galvanic coupling type IBC application on human limb was investigated in both its mathematical model and related experiments. The experimental results showed that the proposed mathematical model was capable in describing the galvanic coupling type IBC under low frequency. Additionally, the calculated result and experimental result also indicated that the electric signal induced by the transmitters of IBC can penetrate deep into human muscle and thus, provide an evident that IBC is capable of acting as networking technique for implantable devices.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bone and Bones / physiology*
  • Computer Simulation
  • Electric Wiring
  • Equipment Design
  • Equipment Failure Analysis
  • Humans
  • Models, Biological*
  • Monitoring, Ambulatory / instrumentation*
  • Pilot Projects
  • Prostheses and Implants*
  • Telemetry / instrumentation*