Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs

J Am Vet Med Assoc. 2009 Dec 1;235(11):1309-13. doi: 10.2460/javma.235.11.1309.

Abstract

Objective: To compare blood glucose concentrations measured with 2 portable blood glucose meters (PBGMs) validated for use in dogs (PBGM-D) and humans (PBGM-H) and an automated chemistry analyzer.

Design: Validation study.

Sample population: 92 samples of fresh whole blood and plasma from 83 dogs with various diseases.

Procedures: Each PBGM was used to measure whole blood glucose concentration, and the automated analyzer was used to measure plasma glucose concentration. Passing-Bablok linear regression and Bland-Altman plots were used to determine correlations and bias between the PBGMs and the automated analyzer. Calculated acceptability limits based on combined inherent instrument imprecision were used with Bland-Altman plots to determine agreement. Clinical relevance was assessed via error grid analysis.

Results: Although correlation between results of both PBGMs and the standard analyzer was > 0.90, disagreement was greater than could be explained by instrument imprecision alone. Mean difference between PBGM-H and chemistry-analyzer values was -15.8 mg/dL. Mean difference between PBGM-D and chemistry-analyzer values was 2.4 mg/dL. Linear regression analysis revealed proportional bias of PBGM-H (greater disagreement at higher glucose concentrations); no proportional bias was detected for PBGM-D. No constant bias was detected for either PBGM. Error grid analysis revealed all measurements from both PBGMs were within zones without an anticipated effect on clinical outcome.

Conclusions and clinical relevance: Neither PBGM had exact agreement with the automated analyzer; however, the disagreement detected did not have serious clinical consequences. Our findings stressed the importance of using the same device for monitoring trends in dogs and using instrument-specific reference ranges.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Automation
  • Blood Chemical Analysis / instrumentation
  • Blood Chemical Analysis / veterinary*
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Dog Diseases / blood*
  • Dogs
  • Humans
  • Retrospective Studies

Substances

  • Blood Glucose