Test-retest variability of intraocular pressure and ocular pulse amplitude for dynamic contour tonometry: a multicentre study

Br J Ophthalmol. 2010 Apr;94(4):419-23. doi: 10.1136/bjo.2009.165142. Epub 2009 Oct 14.

Abstract

Aims: To assess the test-retest variability of intraocular pressure (IOP) and ocular pulse amplitude (OPA) measurements utilising dynamic contour tonometry (DCT) and to evaluate possible influential factors.

Methods: The study included 350 consecutive subjects (175 glaucoma, 175 control; one eye per subject) from seven European centres. IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by DCT (DCT1, DCT2) in a randomised sequence. OPA was also recorded for both DCT measurements. Differences (DCT1-DCT2; OPA1-OPA2; GAT-DCT1; GAT-DCT2) were assessed using the t test. The intraclass coefficient of correlation (ICC) and coefficient of variation (CoV) for DCT and OPA were calculated.

Results: DCT1 was 0.6+/-1.6 mm Hg higher than DCT2 (p<0.001); OPA1 was 0.1+/-0.7 mm Hg higher than OPA2 (p=0.02). Results were not influenced by randomisation test order. In both glaucoma and normal subjects, DCT and OPA showed ICC>0.90 and >0.76, and CoV=4.8-5.0% and 10.3-10.5%, respectively. DCT1 and 2 were 2.4+/-2.6 and 1.8+/-2.6 mm Hg higher respectively than GAT (p<0.001).

Discussion: DCT test-retest variability was almost perfect for IOP and good for OPA. Tonometry measurements with DCT tended to be overestimated compared with GAT.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Female
  • Glaucoma, Open-Angle / physiopathology*
  • Humans
  • Intraocular Pressure / physiology*
  • Male
  • Middle Aged
  • Observer Variation
  • Sensitivity and Specificity
  • Tonometry, Ocular / methods*