Balloon expandable stent implantation for native and recurrent coarctation of the aorta--prospective computed tomography assessment of stent integrity, aneurysm formation and stenosis relief

Heart. 2010 Aug;96(15):1212-6. doi: 10.1136/hrt.2009.170928. Epub 2009 Sep 10.

Abstract

Background: Stenting for aortic coarctation is known to be effective in the medium term. Aneurysm formation following stent implantation is a recognised complication. However, data regarding aortic wall injury and stent integrity following stent placement are sparse.

Objectives: We report comprehensive clinical, echocardiographic and prospective CT follow-up data following stenting for aortic coarctation from a single centre. Methods Full data analysis of all patients undergoing balloon expandable stent implantation and follow-up procedures in a single tertiary congenital cardiac unit.

Results: Between October 2002 and April 2008, we performed 102 coarctation stent procedures on 88 patients. Median age was 20.6 years (range 8.5-65) and median weight 65 kg (range 34-101). 94 stents (26 covered) were implanted. 12 procedures were re-dilatations. Stenting resulted in a reduction of the gradient across the site of coarctation, from a median of 20 mm Hg to 4 mm Hg. There were no procedure-related deaths. Four patients had immediate complications (one requiring emergency surgery). During median follow-up of 34.5 months (range 4.2-72.8), two patients had late complications requiring additional stent procedures. Follow-up CT data are available in 84 patients with MRI in one patient (96.5%). Only one patient developed a procedure-related aortic aneurysm. All stent fractures (n=7) occurred with a single stent design.

Conclusions: Stenting for aortic coarctation and re-coarctation is effective with low immediate complication rates. CT is useful in the longer term for assessment of stent integrity and post-procedural aneurysm formation. Overall incidence of post-procedural aneurysm is rare and stent fractures were not seen with newer generation stents.

Publication types

  • Evaluation Study

MeSH terms

  • Acute Disease
  • Adolescent
  • Adult
  • Aged
  • Aorta / injuries
  • Aortic Aneurysm / diagnostic imaging
  • Aortic Aneurysm / etiology*
  • Aortic Coarctation / therapy*
  • Catheterization / adverse effects
  • Catheterization / methods*
  • Child
  • Equipment Failure
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Recurrence
  • Stents* / adverse effects
  • Tomography, X-Ray Computed
  • Treatment Outcome
  • Young Adult