Safety and effectiveness of the celect inferior vena cava filter: preliminary results

J Vasc Interv Radiol. 2009 Sep;20(9):1188-92. doi: 10.1016/j.jvir.2009.05.033. Epub 2009 Jul 23.

Abstract

Purpose: To evaluate the safety and effectiveness of the Celect inferior vena cava (IVC) filter during implantation, retrieval, and short-term follow-up.

Materials and methods: The clinical data of 73 patients (46 men; age range, 22-89 years) who had a Celect IVC filter implanted between August 2007 and June 2008 were reviewed. Twenty-one (28.8%) presented with pulmonary embolism (PE), 15 (20.54%) with deep vein thrombosis (DVT), 12 (16.4%) with both, and the rest (34.26%) with other symptoms. Indications for filter placement were contraindication to anticoagulation (n = 38; 52%), prophylaxis/added protection (n = 22; 30%), failure of anticoagulation (n = 11; 15%), and complications of anticoagulation (n = 2; 3%). Filters were placed in the infrarenal (n = 71) or suprarenal (n = 2) IVC. Follow-up data were reviewed for filter-related complications and recurrent PE.

Results: All filters were successfully deployed. Immediately after fluoroscopy-guided filter deployment in 61 patients, four filters (6.5%) showed significant tilt. During follow-up (mean, 68 days +/- 73), three patients developed symptoms of PE after filter placement; however, computed tomographic (CT) pulmonary angiography demonstrated new PE in only two. Imaging follow-up with radiography (n = 32), CT (n = 11), and/or angiography (n = 4) in 47 patients (at a mean of 62 days +/- 75) showed no filter migration. Follow-up abdominal CT (at a mean of 69 days +/- 58) was available in 18 patients and demonstrated filter-related problems in seven (39%). These included penetration of filter legs in four and fracture/migration of filter components in one. Fourteen filters were successfully retrieved after a median period of 84 days.

Conclusions: The Celect IVC filter can be safely placed but is related to a high incidence of caval filter leg penetration. Symptomatic PE after filter placement confirmed by CT occurred in 2.8% of patients.

Publication types

  • Clinical Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Equipment Failure Analysis
  • Female
  • Humans
  • Leg Injuries / etiology*
  • Male
  • Middle Aged
  • Pilot Projects
  • Prosthesis Design
  • Treatment Outcome
  • Vena Cava Filters / adverse effects*
  • Wounds, Penetrating / etiology*