An animal study of a newly developed skin-penetrating pad and covering material for catheters to prevent exit-site infection in continuous ambulatory peritoneal dialysis

Artif Organs. 2009 Dec;33(12):1127-33. doi: 10.1111/j.1525-1594.2009.00805.x. Epub 2009 Jul 10.

Abstract

Because currently available peritoneal dialysis catheters are not sufficiently biocompatible with the skin and subcutaneous tissue at the site of penetration, exit-site infection due to pericatheter pocket formation caused by epidermal downgrowth over a long period of time has increasingly become a problem. We developed a new, biocompatible, segmented polyurethane porous material and devised a novel skin-penetrating pad, the form and material of which we optimized for application in peritoneal dialysis catheters. For the extent of tissue ingrowth into this porous material to be examined, test materials with different pore diameters were inserted into hollow silicone tubes and implanted in the subcutaneous tissue of a goat. Four weeks later, the tubes were extracted, and, after the extent of granulation tissue ingrowth was measured, histopathological evaluation was made. Our novel skin-penetrating pad has three disklike layers of the segmented polyurethane material with different pore sizes, into the center of which a polyurethane catheter is inserted. These pads were implanted in the skin of a goat and clinically observed over a 2-year period, after which they were extracted and histopathologically analyzed. In accordance with actual clinical procedures, a commercial CAPD catheter equipped with our skin-penetrating pad was left indwelling in a goat for 4 months, and the performance of the pad was evaluated after repeated periodic infusion and drainage of the dialysate in and out of the abdominal cavity. There was no inflammation of the ingrown tissue in the pores of the segmented polyurethane material as well as the surrounding tissue, which indicated favorable tissue biocompatibility. The extent of tissue ingrowth was greater as the pore size of the material was larger, and the tissue tended to be mature, mainly consisting of collagenous fibers. The skin-penetrating pad using the porous material, of which tissue ingrowth was thus optimized, tightly adhered to the goat skin throughout the 2-year experimental period without any special wound care such as cleansing or disinfection. The performance of the skin-penetrating pad was similarly favorable when attached to a commercial continuous ambulatory peritoneal dialysis catheter. The newly developed segmented polyurethane porous material had excellent tissue biocompatibility and tissue ingrowth. The skin-penetrating pad devised by using this porous material did not cause epidermal downgrowth, suggesting that it may be effective for the prevention of exit-site infection.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Catheter-Related Infections / prevention & control*
  • Catheterization / instrumentation*
  • Catheters, Indwelling
  • Coated Materials, Biocompatible / chemistry*
  • Equipment Design
  • Female
  • Goats
  • Peritoneal Dialysis, Continuous Ambulatory / instrumentation
  • Polyurethanes / chemistry*
  • Porosity
  • Skin / cytology
  • Skin / ultrastructure*

Substances

  • Coated Materials, Biocompatible
  • Polyurethanes