Three dimensional tracking for volumetric spectral-domain optical coherence tomography

Opt Express. 2007 Dec 10;15(25):16808-17. doi: 10.1364/oe.15.016808.

Abstract

We present a three-dimensional (3D) tracker for a clinical ophthalmic spectral domain optical coherence tomography (SD-OCT) system that combines depth-tracking with lateral tracking, providing a stabilized reference frame for 3D data recording and post acquisition analysis. The depth-tracking system is implemented through a real-time dynamic feedback mechanism to compensate for motion artifact in the axial direction. Active monitoring of the retina and adapting the reference arm of the interferometer allowed the whole thickness of the retina to be stabilized to within +/-100 mum. We achieve a relatively constant SNR from image to image by stabilizing the image of the retina with respect to the depth dependent sensitivity of SD-OCT. The depth tracking range of our system is 5.2 mm in air and the depth is adjusted every frame.nhancement in the stability of the images with the depth-tracking algorithm is demonstrated on a healthy volunteer.