Broad angular multilayer analyzer for soft X-rays

Opt Express. 2006 Mar 20;14(6):2533-8. doi: 10.1364/oe.14.002533.

Abstract

Using numerical optimization algorithm, non-periodic Mo/Si, Mo/Be, and Ni/C broad angular multilayer analyzers have been designed. At the wavelength of 13 nm and the angular range of 45~49 degrees , the Mo/Si and Mo/Be multilayer can provide the plateau s-reflectivity of 65% and 45%, respectively. At 5.7 nm, the s-reflectivity of Ni/C multilayer is 16% in the 44~46 degrees range. The non-periodic Mo/Si broad angular multilayer was also fabricated using DC magnetron sputtering, and characterized using the soft X-ray polarimeter at BESSY. The s-reflectivity is higher than 45.6% over the angular range of 45~49 degrees at 13 nm, where, the degree of polarization is more than 99.98%.