Serum levonorgestrel levels and endometrial thickness during extended use of the levonorgestrel-releasing intrauterine system

Contraception. 2009 Jul;80(1):84-9. doi: 10.1016/j.contraception.2009.01.004. Epub 2009 Feb 27.

Abstract

Background: The levonorgestrel-releasing intrauterine system (LNG-IUS) is a contraceptive method approved for five years of use. However, there is some evidence that its life span may be longer. The aim of the study was to evaluate serum levonorgestrel (LNG) and estradiol (E(2)) levels and endometrial thickness every 6 months from 7 to 8 1/2 years after insertion.

Study design: At the end of the approved 5-year life span, no replacement devices were available; therefore, 86 women were allowed to retain the same device for a further 2 years. At the 7-year follow-up visit, the women who consented were again allowed to retain the same device for a further 18 months and were followed-up at six-monthly intervals. At each visit, vaginal sonography was performed to measure endometrial thickness, and a blood sample was taken to measure LNG and E(2).

Results: Eighty-four months after insertion, 67 women aged 34.3+/-0.8 years (mean+/-SEM) (range, 25-49 years) returned for follow-up. Mean+/-SEM LNG levels decreased from an initial 253+/-27 pg/mL (range, 86-760) during the first 2 months following insertion to 137+/-12 (range, 23-393) at 84 months and 119+/-9 pg/mL (range, 110-129) at 102 months of use (+/-SEM). At 84 months of use, mean+/-SEM endometrial thickness was 2.8+/-0.1 mm, increasing to 3.8+/-0.5 mm at 102 months of use. The incidence of amenorrhea decreased from 41.8% at 84 months to 31.5% at 102 months of use. No correlation was found between LNG levels and bleeding patterns; however, a weak correlation was found between high body mass index (kg/m(2)), high weight, and low serum LNG levels. E(2) levels were similar to those of the follicular phase of the menstrual cycle of regularly menstruating women.

Conclusions: During extended use of the LNG-IUS, serum LNG levels were nearly half those found in the first 2 months of use (Wilcoxon signed rank test); serum E(2) levels were normal. Despite the very thin endometrium, menstrual bleeding was reinstated in many cases. At the end of its 5-year life span, there is a window for changing the LNG-IUS, and physicians and users should not be concerned about delaying replacement of the device for a short time beyond the approved life span; however, maintaining the same device long after its approved life span cannot be recommended.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Contraceptive Agents, Female / administration & dosage
  • Contraceptive Agents, Female / blood*
  • Endometrium / drug effects*
  • Estradiol / administration & dosage
  • Estradiol / blood*
  • Female
  • Humans
  • Intrauterine Devices, Medicated*
  • Levonorgestrel / administration & dosage
  • Levonorgestrel / blood*
  • Middle Aged
  • Time Factors

Substances

  • Contraceptive Agents, Female
  • Estradiol
  • Levonorgestrel