[Stapes surgery : first experiences with the new Soft-CliP piston]

HNO. 2009 May;57(5):509-13. doi: 10.1007/s00106-009-1899-y.
[Article in German]

Abstract

Background: The first hearing results with a new stapes prosthesis with clip function (Soft-CliP(R) piston) are presented.

Patients and methods: This new prosthesis was used in 15 patients (mean age 45.2 years; range 21-63 years) undergoing routine stapes surgery. Soft-CliP(R) piston prostheses with a shaft diameter of 0.4 mm and a length ranging from 4.25 mm to 5.5 mm were used. Postoperative audiological testing and measurement of the air-bone gap were performed after an average of 47.3 days and compared with the preoperative values.

Results: The median observed postoperative air-bone gap (ABG) was 8.33 dB +/-4.16 dB. All patients had less than 20 dB ABG and in 53.3% of cases was less than 10 dB. The operating time showed a clear difference between the left (66.5 min +/-37.79 min) and right ears (47.2 min +/-11.08 min).

Discussion: This new prosthesis design greatly facilitates a very difficult step in stapes surgery, the prosthesis fixation to the incus. The first postoperative hearing results are very promising but long-term results in a larger group of patients are still pending.

Publication types

  • Clinical Trial
  • English Abstract

MeSH terms

  • Adult
  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pilot Projects
  • Stapes Surgery / instrumentation*
  • Stapes Surgery / methods*
  • Surgical Instruments*
  • Treatment Outcome