Image correction during large and rapid B(0) variations in an open MRI system with permanent magnets using navigator echoes and phase compensation

Magn Reson Imaging. 2009 Sep;27(7):988-93. doi: 10.1016/j.mri.2009.01.022. Epub 2009 Apr 15.

Abstract

An open permanent magnet system with vertical B(0) field and without self-shielding can be quite susceptible to perturbations from external magnetic sources. B(0) variation in such a system located close to a subway station was measured to be greater than 0.7 microT by both MRI and a fluxgate magnetometer. This B(0) variation caused image artifacts. A navigator echo approach that monitored and compensated the view-to-view variation in magnetic resonance signal phase was developed to correct for image artifacts. Human brain imaging experiments using a multislice gradient-echo sequence demonstrated that the ghosting and blurring artifacts associated with B(0) variations were effectively removed using the navigator method.