Comparison of antibody monitoring system with flow cytometric crossmatch test in renal transplant recipients with high panel-reactive antibody

Nephron Clin Pract. 2009;111(4):c260-4. doi: 10.1159/000209153. Epub 2009 Mar 17.

Abstract

Background/aims: The antibody monitoring system (AMS) is a recently developed enzyme-linked immunosorbent assay (ELISA) crossmatch assay to detect donor-specific anti-HLA immunoglobulin G antibodies (DS-HLA Abs). This study was conducted to compare the AMS with the flow cytometric crossmatch (FCXM) test in renal transplant recipients with high panel-reactive antibody (PRA).

Methods: Thirty-two sera were obtained from 10 patients with panel reactivity above 50%. When anti-HLA Ab was detected by ELISA PRA and the matched donor had the corresponding HLA antigen, it was considered to indicate DS-HLA Ab. The results of the AMS assay and FCXM were compared with the DS-HLA Abs.

Results: Twenty-three (71.9%) sera were positive for DS-HLA Abs by ELISA PRA. The AMS assay showed that the number of compatible sera with DS-HLA Abs was 27 (84.4%), and it was significantly concordant (kappa = 0.649, p < 0.0001). For FCXM, the number of compatible sera with DS-HLA Abs was 26 (81.3%), and it was also significantly concordant (kappa = 0.614, p < 0.0001). There was a significant degree of concordance between the AMS assay and FCXM in detection of DS-HLA Abs (kappa = 0.452, p = 0.010).

Conclusion: The AMS assay is comparable to FCXM in detecting DS-HLA Abs in high PRA recipients.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Enzyme-Linked Immunosorbent Assay / methods*
  • Flow Cytometry / methods*
  • HLA-D Antigens / immunology*
  • Humans
  • Immunoglobulin G / blood*
  • Kidney Transplantation / immunology*
  • Monitoring, Immunologic / methods*