This work aims at comparing the delta(13)C(VPDB) of mono-aromatic hydrocarbons benzene, toluene, ethylbenzene and xylene isomers (BTEX) measured by elemental analyser (EA)-isotope ratio mass spectrometer (IRMS) with the delta(13)C(VPDB) measured on the same compounds by headspace solid phase microextraction - GC/C-IRMS (hSPME - GC/C-IRMS) with the final goal of using these compounds as internal standards on the latter system. The EA-IRMS measurements were done using calcium and lithium carbonate isotopic reference materials: NBS19 and L-SVEC for establishing the delta(13)C(VPDB) scale. The EA-IRMS measurements with helium dilution of a set of five reference materials (USGS40, USGS41, IAEA-CH-6, IAEA-CH-3 and IAEA-601) show systematic bias of 1 per thousand relative to their assigned values. This bias due to the dilution mechanism in the used ConfloII interface device could not be avoided. As the selected hydrocarbons: BTEX could not be analysed by EA-IRMS without helium dilution, their delta(13)C(VPDB) must be corrected from this observed bias using an external calibration. The CO(2) gas calibrated using EA-IRMS without helium dilution, was used as an in-house reference for the delta(13)C(VPDB) measurements of the BTEX by the hSPME - GC/C-IRMS system. The comparison made between the delta(13)C(VPDB) measured on the same BTEX compounds by EA-IRMS (with external calibration) and by hSPME - GC/C-IRMS techniques showed good agreement.