Evaluation of internal standards for the analysis of ignitable liquids in fire debris

J Forensic Sci. 2009 Mar;54(2):320-7. doi: 10.1111/j.1556-4029.2008.00954.x. Epub 2009 Jan 27.

Abstract

An evaluation of eight compounds for use as an internal standard in fire debris analysis was conducted. Tests were conducted on tetrachloroethylene, chlorobenzene, n-octylbenzene, 3-phenyltolune, and deuterated compounds toluene-d8, styrene-d8, naphthalene-d8, and diphenyl-d10 to measure the extraction efficiency of each compound in the presence of an interfering volatile compound (carbon disulfide). Other tests were conducted to evaluate whether or not the presence of an ignitable liquid or pyrolysis/combustion products from fire debris would interfere with the identification of these compounds when used as an internal standard. The results showed that while any of the eight compounds could be used as an internal standard in fire debris analysis, the more volatile compounds (toluene-d8, tetrachloroethylene, chlorobenzene, and styrene-d8) showed better extraction efficiencies at room temperature than when heated to 60 degrees C. Each of the less volatile compounds (naphthalene-d8, diphenyl-d10, n-octylbenzene, and 3-phenyltolune) performed well during extraction at 60 degrees C, while naphthalene-d8 showed better extraction efficiency in the presence of competing volatiles when extracted at room temperature.