Consistency of apex locator function: a clinical study

J Endod. 2009 Feb;35(2):179-81. doi: 10.1016/j.joen.2008.10.017. Epub 2008 Dec 12.

Abstract

The consistency of apex locators was determined by calculating the dysfunction frequency. Electronic working length (EWL) was determined in 507 patients requiring endodontic treatment. Different clinical parameters were recorded including tooth vitality, presence of obliteration, and metallic restoration. Two apex locators were used (Root ZX [Morita, Tokyo, Japan] and Raypex5 [VDW, Munich, Germany]). Apex locator performance was considered "consistent" when the scale bars were stable and moved only in correspondence to the movement of file in the root canal. A working length radiograph with files set to the EWL was performed. EWL were considered "acceptable" when the file tip was located 0 to 2 mm short of the radiographic apex. The function of apex locators was consistent in 85% of the patients (429/507 [99% confidence interval, 80-88]). The inconsistent measurements were strongly associated with partially or totally obliterated root canals (p < 0.0001). Radiographically, 97% of consistent measurements were "acceptable."

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Electrical Equipment and Supplies*
  • Female
  • Humans
  • Logistic Models
  • Male
  • Middle Aged
  • Odontometry / instrumentation*
  • Tooth Apex / anatomy & histology*
  • Young Adult