Transcranial shear-mode ultrasound: assessment of imaging performance and excitation techniques

IEEE Trans Med Imaging. 2009 May;28(5):763-74. doi: 10.1109/TMI.2008.2012017. Epub 2009 Jan 13.

Abstract

Transcranial ultrasound imaging is limited by poor acoustic windows and skull induced distortions to the beam. Shear waves in the skull have a better impedance match with longitudinal waves in water and thereby produce a more coherent focus inside the skull. This study presents work on an imaging technique that utilizes shear-wave propagation through the skull. The pulse-echo lateral distortion introduced by the skull was analyzed by imaging a point scatterer behind ex vivo human craniums at 1 MHz. Brightness images of the target obtained with either shear-mode or conventional longitudinal-mode transmission in the bone were assessed to quantify lateral resolution. As compared to longitudinal-mode transmission, it was found that the use of shear-mode resulted in improved localization along the propagation (depth) axis at the expense of degraded lateral resolution. The signal-to-noise ratio (SNR) limitations introduced by severe attenuation of shear-waves in the skull were overcome with frequency modulated (FM) coded excitations. This gain in SNR was exchanged with resolution and used for compensation of frequency-dependent attenuation in the skull, resulting in a greater than 20% improvement in lateral resolution for both modes of transcranial transmission. The results are an important step towards enhancing the quality of transcranial sonography.

MeSH terms

  • Algorithms
  • Artifacts
  • Equipment Design
  • Fourier Analysis
  • Humans
  • Image Enhancement / methods*
  • Signal Processing, Computer-Assisted*
  • Skull / anatomy & histology
  • Ultrasonography, Doppler, Transcranial / methods*