Pannus-related prosthetic valve dysfunction and life-threatening aortic regurgitation

J Heart Valve Dis. 2008 Nov;17(6):666-9.

Abstract

Prosthetic valve dysfunction is a rare but life-threatening condition. A 66-year-old woman presented with shock 15 years after aortic valve replacement with a tilting-disc valve. Imaging demonstrated severe aortic insufficiency and a fixed-open prosthetic valve. Reoperation revealed pannus ingrowth from the aortic aspect, resulting in immobility of the occluder. A bioprosthetic valve was installed and the patient recovered uneventfully. The diagnosis and surgical management of this problem are discussed.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Aortic Valve / pathology*
  • Aortic Valve / surgery
  • Aortic Valve Insufficiency / diagnostic imaging
  • Aortic Valve Insufficiency / etiology*
  • Aortic Valve Insufficiency / surgery
  • Aortic Valve Stenosis / surgery
  • Bioprosthesis
  • Female
  • Fibrosis
  • Heart Valve Prosthesis / adverse effects*
  • Humans
  • Prosthesis Failure*
  • Reoperation
  • Shock, Cardiogenic / etiology
  • Shock, Cardiogenic / surgery
  • Thrombosis / etiology
  • Thrombosis / surgery
  • Ultrasonography