Triclosan determination in water related to wastewater treatment

Talanta. 2007 Jul 31;72(5):1650-4. doi: 10.1016/j.talanta.2007.03.024. Epub 2007 Mar 23.

Abstract

Triclosan in the waste, river and sea water samples collected in Hong Kong was analyzed by using gas chromatography-ion trap mass spectrometry method. (13)C(12)-triclosan was used as internal standard for the quantitative analysis. Water samples were prepared and cleaned-up by using a C(18) solid-phase extraction cartridge. The recoveries of triclosan in spiked coastal water at three different concentrations ranged from 83 to 110%. The method detection limit was 0.25 ng/L for triclosan in 1-L water and the relative standard deviations and relative error were less than 11.0 and 12.3%, respectively (n=3). The method was successfully applied to analyze water samples collected from rivers, coastal water bodies and wastewater treatment plants at ng/L levels.