Sequential injection analysis system for on-line monitoring of l-cysteine concentration in biological processes

Talanta. 2005 Dec 15;68(2):187-92. doi: 10.1016/j.talanta.2005.06.049. Epub 2005 Jul 28.

Abstract

A sequential injection analysis (SIA) system was developed to monitor the concentration of l-cysteine in biological processes on-line. It is based on the redox reaction of l-cysteine with iron(III) in the presence of 1,10-phenanthroline (phen) and the detection of the red-iron(II)-phen complex with a spectrophotometry. The system was fully automated using software written in the LabVIEWtrade mark development environment. A number of system variables such as the flow rate of the carrier buffer solution, the volume ratio of the sample to the reagents, and the reaction coil length, etc., were evaluated to increase the sensitivity and performance of the SIA system. Under partially optimized operating conditions the performance of the SIA system was linear up to a concentration of l-cysteine of 1mM (R(2)=0.998) with a detection limit of 0.005mM and a sample frequency of 15hr(-1). The SIA system was employed to monitor the concentration of l-cysteine on-line in a continuously stirred reactor and a fermentation process of Saccharomyces cerevisiae. The on-line monitored data were in good agreement with the off-line data measured by a HPLC with a fluorescence detector (n=15, R(2)=09899).