Manganese detection in marine sediments: anodic vs. cathodic stripping voltammetry

Talanta. 2005 Jan 30;65(2):423-9. doi: 10.1016/j.talanta.2004.06.038.

Abstract

Three different electroanalytical techniques for the detection of manganese in marine sediments are evaluated. The anodic stripping voltammetry of manganese at an in situ bismuth-film-modified boron-doped diamond electrode and cathodic stripping voltammetry at a carbon paste electrode are shown to lack the required sensitivity and reproducibility whereas cathodic stripping voltammetry at a bare boron-doped diamond electrode is shown to be reliable and selective with a limit of detection, from applying a 60s accumulation period of 7.4 x 10(-7)M and a sensitivity of 0.24AM(-1). The method was used to evaluate the manganese content of marine sediments taken from Sibenik, Croatia.