Lead-selective poly(vinyl chloride) membrane electrodes based on acyclic dibenzopolyether diamides

Talanta. 1997 Jun;44(6):1131-5. doi: 10.1016/s0039-9140(97)00021-0.

Abstract

Lipophilic acyclic dibenzopolyether diamides, 12 kinds, have been designed to prepare solvent polymeric membrane ion-selective electrodes (ISEs) for Pb(2+). The ionophores include 1,5-bis[2-(N,N-dialkylcarbamoylmethoxy)phenoxy]-3-oxapentanes1-4, 1,5-bis[2-(N,N-dialkylcarbamoylpentadecyloxy)phenoxy]-3-oxapentanes 5-8, and 1,2-bis[2-(2'-N,N-dialkylcarbamoylpentadecyloxy)phenoxy]ethanes 9-12. Linear response concentration range of the ISE based on 9 is 3 x 10(-2) - 1 x 10(-6) M of Pb(2+) (average slope = 28.5 mV decade(-1)). Potentiometric selectivities of the ISEs based on 1-12 for Pb(2+) over other heavy metal cations, alkali metal cations, and alkaline earth metal cations have been assessed. These ISEs exhibit remarkably high selectivities for Pb(2+) relative to heavy metal cations, such as Cu(2+), Fe(2+), and Ni(2+), the selectivity coefficients (K(Pot)(Pb,Cu)) being 5 x 10(-5) - 6 x 10(-5) for 1-4 and ca. 6 x 10(-4) for 9. For the Pb(2+) selectivities over alkali metal cations, such as Na(+) and K(+), 9 which has an ethylene glycol spacer and a N,N-diethyl group is superior to other dibenzopolyether diamide ionophores 1-8 and 10-12.