Empirical dead-time corrections for synchrotron sources

J Synchrotron Radiat. 2008 Nov;15(Pt 6):612-7. doi: 10.1107/S0909049508022358. Epub 2008 Oct 3.

Abstract

An experimental comparison of models for performing dead-time corrections of photon-counting detectors at synchrotron sources is presented. The performance of several detectors in the three operating modes of the Advanced Photon Source is systematically compared, with particular emphasis on asymmetric fill patterns. Several simple and well known correction formulas are evaluated. The results demonstrate the critical importance of detector speed and synchrotron fill pattern in selecting the proper dead-time correction.