Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard

Clin Chem Lab Med. 2008;46(10):1450-7. doi: 10.1515/CCLM.2008.304.

Abstract

Background: Ferritin standardisation is problematical due to the heterogeneity of ferritin isoforms and the antibodies used in its immunoassay, and the lack of a reference measurement procedure. We investigated the performance of the 1st (liver), 2nd (spleen) and 3rd (recombinant) International Standards (ISs) for ferritin in major assays.

Methods: The ferritin in a serum pool 'spiked' with either the 2nd or 3rd IS for ferritin was measured by 52 laboratories using five automated methods and the recovery of the target values calculated. A smaller serum pool was 'spiked' with the 1st IS for a limited recovery exercise. The ferritin values of five serum samples were also measured and recalculated relative to the ISs.

Results: Recoveries of each of the 2nd and 3rd ISs were 90%-110% for four of five methods; recoveries of the 1st IS were 104% and 111% for two of three methods claiming traceability to this IS. One method significantly over-recovered each of the IS (124%-155%). Recalculating the ferritin values of the serum samples relative to the IS reduced the overall inter-method agreement, largely because of the anomalous over-recovery of the IS by one method.

Conclusions: The use of the 3rd IS to standardise assays will minimise assay drift due to manufacturers adopting a 'harmonisation' approach in which the calibration is adjusted to conform to overall mean values. Standardisation against the current IS also ensures compliance with the European Union In-Vitro Diagnostic Directive which requires traceability of assay calibrators to reference materials of a higher order. Assay drift may result in poor sensitivity and specificity in the diagnosis of iron status, and would require laboratories to continually re-evaluate reference intervals.

MeSH terms

  • Automation
  • Ferritins / analysis*
  • Humans
  • Immunoassay / methods*
  • Immunoassay / standards*
  • Internationality
  • Reference Standards
  • Reproducibility of Results

Substances

  • Ferritins