Prospective, randomized comparison of cobalt-chrome and titanium trilock femoral stems

J Arthroplasty. 2009 Sep;24(6):831-6. doi: 10.1016/j.arth.2008.06.035. Epub 2008 Oct 5.

Abstract

The purpose of this article was to demonstrate the efficacy of a cementless, flat, tapered wedge femoral stem and compare cobalt-chrome and titanium femoral stems with this design. Three hundred ninety femoral stems observed for a mean of 4.7 years (2.0-8.9 years) were prospectively evaluated with clinical and radiographic follow-up. Hips were stratified by Dorr classification, bone stock (femoral index), size of implant used, and material of femoral implant. Survivorship of the femoral stem at 8.9 years was 99.8% with no significant difference between cohorts. Thigh pain (4.9%) was more common with cobalt-chrome femoral stems (6.5%) than titanium femoral stems (3.1%). The flat, tapered wedge femoral stem design provides excellent femoral reconstruction in total hip arthroplasty. On the basis of this study, we use titanium femoral stems.