Micro insert: a prototype full-ring PET device for improving the image resolution of a small-animal PET scanner

J Nucl Med. 2008 Oct;49(10):1668-76. doi: 10.2967/jnumed.107.050070. Epub 2008 Sep 15.

Abstract

A full-ring PET insert device should be able to enhance the image resolution of existing small-animal PET scanners.

Methods: The device consists of 18 high-resolution PET detectors in a cylindric enclosure. Each detector contains a cerium-doped lutetium oxyorthosilicate array (12 x 12 crystals, 0.72 x 1.51 x 3.75 mm each) coupled to a position-sensitive photomultiplier tube via an optical fiber bundle made of 8 x 16 square multiclad fibers. Signals from the insert detectors are connected to the scanner through the electronics of the disabled first ring of detectors, which permits coincidence detection between the 2 systems. Energy resolution of a detector was measured using a (68)Ge point source, and a calibrated (68)Ge point source stepped across the axial field of view (FOV) provided the sensitivity profile of the system. A (22)Na point source imaged at different offsets from the center characterized the in-plane resolution of the insert system. Imaging was then performed with a Derenzo phantom filled with 19.5 MBq of (18)F-fluoride and imaged for 2 h; a 24.3-g mouse injected with 129.5 MBq of (18)F-fluoride and imaged in 5 bed positions at 3.5 h after injection; and a 22.8-g mouse injected with 14.3 MBq of (18)F-FDG and imaged for 2 h with electrocardiogram gating.

Results: The energy resolution of a typical detector module at 511 keV is 19.0% +/- 3.1%. The peak sensitivity of the system is approximately 2.67%. The image resolution of the system ranges from 1.0- to 1.8-mm full width at half maximum near the center of the FOV, depending on the type of coincidence events used for image reconstruction. Derenzo phantom and mouse bone images showed significant improvement in transaxial image resolution using the insert device. Mouse heart images demonstrated the gated imaging capability of the device.

Conclusion: We have built a prototype full-ring insert device for a small-animal PET scanner to provide higher-resolution PET images within a reduced imaging FOV. Development of additional correction techniques are needed to achieve quantitative imaging with such an insert.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Animals
  • Diagnostic Imaging / instrumentation
  • Diagnostic Imaging / methods
  • Electrocardiography / methods
  • Equipment Design
  • Fluorine Radioisotopes / pharmacology
  • Fluorodeoxyglucose F18 / pharmacology
  • Image Processing, Computer-Assisted
  • Mice
  • Nuclear Medicine / methods
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation*
  • Positron-Emission Tomography / methods*

Substances

  • Fluorine Radioisotopes
  • Fluorodeoxyglucose F18