Spectromicroscopic film dosimetry for high-energy microbeam from synchrotron radiation

Appl Radiat Isot. 2009 Jan;67(1):155-9. doi: 10.1016/j.apradiso.2008.08.002. Epub 2008 Aug 6.

Abstract

A microscope with band-pass filters was used to measure the optical-density distribution of GafChromic films irradiated with multi-slit microbeam X-rays. The planar width was 25 microm, and the center-to-center distance was 200 microm. The peak and valley dose rates in air were found to be 120 and 0.7Gy/s, respectively. In a polymethylmethacrylate phantom, the peak-to-valley dose ratio decreased to 80 at a 1-mm depth. Doses calculated with the PENELOPE code agreed with those around the peak but became smaller in the valley.

MeSH terms

  • Film Dosimetry / methods*
  • Microscopy / methods
  • Synchrotrons*