A new approach to the application of solid phase extraction disks with LC-MS/MS for the analysis of drugs on a 96-well plate format

J Pharm Biomed Anal. 2009 Nov 1;50(4):556-62. doi: 10.1016/j.jpba.2008.07.014. Epub 2008 Jul 29.

Abstract

A new 96-well disk solid phase extraction sample preparation technique which does not involve vacuum pumps integrated with liquid chromatographic tandem mass spectrometric (LC-MS/MS) was developed for high throughput determination of benzodiazepines (nordiazepam, diazepam, lorazepam and oxazepam). In addition, the method completely allows the re-use of the SPE disk membranes for subsequent analyses after re-conditioning. The method utilizes a robotic autosampler for parallel extractions in a 96-well plate format. Results have been presented for independent extractions from three matrices; phosphate buffer solution, urine, and plasma. Factors affecting data reproducibility, extraction kinetics, sample throughput, and reliability of the system were investigated and optimized. A total time required per sample was 0.94 min using 96-well format. Method reproducibility was < or =9% relative standard deviation for all three matrices. Limits of detection and quantitation recorded were respectively in the range 0.02-0.15 and 0.2-2.0 ng/mL with linearity ranging from 0.2 to 500 ng/mL for all matrices.

MeSH terms

  • Automation
  • Benzodiazepines / analysis*
  • Chromatography, Liquid / methods*
  • Humans
  • Reproducibility of Results
  • Solid Phase Extraction / methods
  • Tandem Mass Spectrometry / methods*
  • Time Factors

Substances

  • Benzodiazepines