A practical alternative to chemiluminescence-based detection of nitrogen dioxide: cavity attenuated phase shift spectroscopy

Environ Sci Technol. 2008 Aug 15;42(16):6040-5. doi: 10.1021/es703204j.

Abstract

We present results obtained from a greatly improved version of a previously reported nitrogen dioxide monitor (Anal Chem. 2005, 77, 724-728) that utilizes cavity attenuated phase shift spectroscopy (CAPS). The sensor, which detects the optical absorption of nitrogen dioxide within a 20 nm bandpass centered at 440 nm, comprises a blue light emitting diode, an enclosed stainless steel measurement cell (26 cm length) incorporating a resonant optical cavity of near-confocal design and a vacuum photodiode detector. An analog heterodyne detection scheme is used to measure the phase shift in the waveform of the modulated light transmitted through the cell induced by the presence of nitrogen dioxide within the cell. The sensor, which operates at atmospheric pressure, fits into a 19 in.-rack-mounted instrumentation box, weighs 10 kg, and utilizes 70 W of electrical power with pump included. The sensor response to nitrogen dioxide (calculated as the cotangent of the phase shift) is demonstrated to be linear (r2 > 0.9999) within +/- 1 ppb over a range of 0-320 ppb (by volume). The device exhibits a detection limit (3sigma precision) of less than 60 parts per trillion (0.060 ppb) with 10 s integration, a value derived from measurements at NO2 concentration levels of both 0 and 20 ppb; the detection limit improves as the integration time is increased to several hundred seconds. The observed baseline drift is less than +/- 0.5 ppb overthe course of a month. An intercomparison of measurements of ambient NO2 concentrations over several days using this sensor with a quantum cascade laser-based infrared absorption spectrometer and a standard chemiluminescence-based NOx analyzer is presented. The data from the CAPS sensor are highly correlated (r2 > 0.99) with the other two instruments. The absolute agreement between the CAPS and each of the two other instruments is within the expected statistical noise associated with the infrared laser-based absorption spectrometer (+/- 0.3 ppb with 10 s sampling) and chemiluminescence analyzer (+/- 0.4 ppb with 60 s averaging). The major limitation concerning accuracy is a direct spectral interference with phototchemically produced 1,2-dicarbonyl species (e.g., glyoxal, methylglyoxal). However, this interference can be readily removed by shifting the detection band to a slightly longer wavelength and ensuring that the lower edge of the detection band is greater than 455 nm.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Air Pollutants / chemistry*
  • Environmental Monitoring / instrumentation*
  • Environmental Monitoring / methods
  • Luminescence
  • Nitrogen Dioxide / chemistry*
  • Spectrum Analysis / instrumentation*
  • Spectrum Analysis / methods

Substances

  • Air Pollutants
  • Nitrogen Dioxide