Contact stress analysis of the conforming post-cam mechanism in posterior-stabilized total knee arthroplasty

J Arthroplasty. 2008 Aug;23(5):736-43. doi: 10.1016/j.arth.2007.05.023.

Abstract

The present study evaluated the effects of extent of conformity of post-cam design on contact area and stress at post-cam mechanism using 4 different posterior-stabilized prostheses. TRAC and Alpina with full-conformed post-cams exhibited the largest contact area at 90 degrees and 120 degrees . PFC sigma RPF with partial conformed post-cam had the largest contact area at 150 degrees . Scorpio NRG with less conformed post-cam had smaller contact area than the others. Lifting of femoral component decreased contact area and increased contact stress of TRAC and Alpina. Recent modifications of post-cam design have increased contact area, contributing to lower contact stress. None of these prostheses exhibited constant low contact stress throughout flexion. Further modifications of post-cam mechanism are necessary to provide lower contact stress throughout deep knee flexion.

Publication types

  • Comparative Study

MeSH terms

  • Arthroplasty, Replacement, Knee / instrumentation*
  • Biomechanical Phenomena
  • Computer-Aided Design*
  • Femur / surgery
  • Humans
  • Knee Prosthesis*
  • Materials Testing / methods*
  • Polyethylene
  • Prosthesis Design*
  • Range of Motion, Articular
  • Stress, Mechanical

Substances

  • Polyethylene