Renal cyst pseudoenhancement at multidetector CT: what are the effects of number of detectors and peak tube voltage?

Radiology. 2008 Sep;248(3):910-6. doi: 10.1148/radiol.2482071583. Epub 2008 Jul 15.

Abstract

Purpose: To determine the effect of the number of detectors and peak tube voltage on renal cyst pseudoenhancement in a phantom model.

Materials and methods: This study on computed tomographic (CT) phantoms did not require institutional review board approval. The renal compartments of a CT phantom were filled with iodinated contrast material diluted to attain attenuations of 40, 140, and 240 HU. Saline-filled cylinders simulating cysts of varying diameters (range, 0.7-3.0 cm) were serially suspended in the renal compartments and scanned at 80, 90, 100, 120, and 140 kVp in 16-detector (n = 3) and 64-detector (n = 2) CT scanners. Generalized estimating equations were used to determine predictors of cyst pseudoenhancement (defined as a >10 HU increase in cyst attenuation when the background renal attenuation increased from 40 to 140 or 240 HU).

Results: Pseudoenhancement was seen with higher frequency (59 [61%] of 96 cysts vs 52 [39%] of 132 cysts, P < .05) and magnitude (17 vs 13 HU, P < .005) with 64- rather than with 16-detector scanners. Pseudoenhancement was also seen with higher frequency (25 [42%] of 60 cysts vs 11 [18%] of 60 cysts, P < .005) and magnitude (18 vs 13 HU, P < .05) at 140 kVp than at 80 or 90 kVp. Cyst pseudoenhancement increased with higher background renal enhancement (P < .005) and smaller cyst diameter (P < .05). The number of detectors, peak tube voltage, renal parenchymal enhancement level, and cyst diameter were independent predictors of cyst pseudoenhancement.

Conclusion: Lower tube voltage settings may be useful when accurate differentiation between small renal cysts and solid masses is critical, particularly for 64-detector CT scanners.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, N.I.H., Extramural

MeSH terms

  • Humans
  • Kidney / diagnostic imaging*
  • Kidney Diseases, Cystic / diagnostic imaging*
  • Phantoms, Imaging
  • Radiographic Image Enhancement / instrumentation*
  • Radiographic Image Enhancement / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / instrumentation*
  • Tomography, X-Ray Computed / methods*