Disposable mercury-free cell-on-a-chip devices with integrated microfabricated electrodes for the determination of trace nickel(II) by adsorptive stripping voltammetry

Anal Chim Acta. 2008 Aug 1;622(1-2):111-8. doi: 10.1016/j.aca.2008.05.051. Epub 2008 May 28.

Abstract

This work reports the fabrication of disposable three-electrode cells with integrated sputtered metal-film electrodes. The working electrode was a bismuth-film electrode (BiFE) while the reference and counter electrodes were made of Ag and Pt, respectively. The deposition of the metal layers was carried out by sputtering of the respective metals on a silicon substrate while the exact geometry of the electrodes was defined via a metal mask placed on the substrate during the deposition process. Initially, the electrodes were characterised by cyclic voltammetry. The utility of these devices was tested for the trace determination of Ni(II) by square wave adsorptive stripping voltammetry (SWAdSV) after complexation with dimethylglyoxime (DMG). The experimental variables (the presence of oxygen, the DMG concentration, the preconcentration potential, the accumulation time and the SW parameters), as well as potential interferences, were investigated. Using the selected conditions, the 3sigma limit of detection was 100 ng L(-1) for Ni(II) (for 90 s of preconcentration) and the relative standard deviation for Ni(II) was 2.3% at the 10 microg L(-1) level (n=8). Finally, the method was applied to the determination of Ni(II) in a certified river water sample.