High-performance liquid chromatographic determination of propylthiouracil in human urine by post-column iodine-azide reaction

J Pharm Biomed Anal. 2008 Sep 10;48(1):145-50. doi: 10.1016/j.jpba.2008.04.031. Epub 2008 May 14.

Abstract

A sensitive, selective and simple post-column detection method for the determination of propylthiouracil (PTU) based on its sensitizing induction on iodine-azide reaction and the combination technique of high-performance liquid chromatography has been presented. The analysis was conducted in the optimum conditions for iodine-azide detection system and HPLC separation. The linear range, the lower limit of detection and quantification for PTU in urine were established at the levels of 0.4-1.0 nmol/ml urine, 0.3 nmol/ml urine and 0.4 nmol/ml urine, respectively.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antithyroid Agents / urine*
  • Azides / chemistry*
  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Female
  • Humans
  • Iodine / chemistry*
  • Middle Aged
  • Propylthiouracil / urine*
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Antithyroid Agents
  • Azides
  • Propylthiouracil
  • Iodine