Determination of d-limonene in adipose tissue by gas chromatography-mass spectrometry

J Chromatogr B Analyt Technol Biomed Life Sci. 2008 Jul 1;870(1):68-73. doi: 10.1016/j.jchromb.2008.06.002. Epub 2008 Jun 6.

Abstract

We developed a novel method for analyzing d-limonene levels in adipose tissue. Fat samples were subjected to saponification followed by solvent extraction. d-Limonene in the sample extract was analyzed using gas chromatography-mass spectrometry (GC-MS) with selected ion monitoring. Linear calibration curves were established over the mass range of 79.0-2529 ng d-limonene per 0.1g of adipose tissue. Satisfactory within-day precision (R.S.D. 6.7-9.6%) and accuracy (%difference of -2.7 to 3.8%) and between-day precision (R.S.D. 6.0-10.7%) and accuracy (%difference of 1.8-2.6%) were achieved. The assay was successfully applied to human fat biopsy samples from a d-limonene feeding trial.

Publication types

  • Evaluation Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adipose Tissue / chemistry*
  • Calibration
  • Cyclohexenes / chemistry*
  • Gas Chromatography-Mass Spectrometry / methods*
  • Humans
  • Limonene
  • Terpenes / chemistry*

Substances

  • Cyclohexenes
  • Terpenes
  • Limonene