Standardization of C-peptide measurements

Clin Chem. 2008 Jun;54(6):1023-6. doi: 10.1373/clinchem.2007.101287. Epub 2008 Apr 17.

Abstract

Background: C-peptide is a marker of insulin secretion in diabetic patients. We assessed within- and between-laboratory imprecision of C-peptide assays and determined whether serum calibrators with values assigned by mass spectrometry could be used to harmonize C-peptide results.

Methods: We sent 40 different serum samples to 15 laboratories, which used 9 different routine C-peptide assay methods. We also sent matched plasma samples to another laboratory for C-peptide analysis with a reference mass spectrometry method. Each laboratory analyzed 8 of these samples in duplicate on each of 4 days to evaluate within- and between-day imprecision. The same 8 samples were also used to normalize the results for the remaining samples to the mass spectrometry reference method.

Results: Within- and between-run CVs ranged from <2% to >10% and from <2% to >18%, respectively. Normalizing the results with serum samples significantly improved the comparability among laboratories and methods. After normalization, the differences among laboratories in mean response were no longer statistically significant (P = 0.24), with least-squares means of 0.93-1.02.

Conclusions: C-peptide results generated by different methods and laboratories do not always agree, especially at higher C-peptide concentrations. Within-laboratory imprecision also varied, with some methods giving much more consistent results than others. These data show that calibrating C-peptide measurement to a reference method can increase comparability between laboratories.

Publication types

  • Multicenter Study
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • C-Peptide / blood*
  • Calibration
  • Chromatography, Liquid
  • Humans
  • Immunoassay / methods
  • Least-Squares Analysis
  • Mass Spectrometry / standards
  • Reference Standards

Substances

  • C-Peptide