Estimation of optical properties of turbid media: experimental comparison of spatially and temporally resolved reflectance methods

Appl Opt. 2008 Apr 10;47(11):1734-9. doi: 10.1364/ao.47.001734.

Abstract

We compare two methods for the optical characterization of turbid media. The estimates of the absorption and reduced scattering coefficients (mu(a) and mu(')(s)) by a spatially resolved method and a time-resolved method are performed on tissue-like phantoms. Aqueous suspension of microspheres and Intralipid are used as scattering media with the addition of ink as an absorber. mu(')(s) is first measured on weakly absorbing media. The robustness of these measurements is then tested with respect to a variation of mu(a). The spatially resolved method gave more accurate estimates for mu(')(s) whereas the time-resolved method gave better results for mu(a) estimates.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Algorithms*
  • Colloids / chemistry*
  • Nephelometry and Turbidimetry / methods*
  • Photometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Colloids