Atmospheric ozone measurement with an inexpensive and fully automated porous tube collector-colorimeter

Talanta. 2008 Jan 15;74(4):958-64. doi: 10.1016/j.talanta.2007.08.002. Epub 2007 Aug 9.

Abstract

The bleaching action of ozone on indigo and related compounds is well known. We describe sensitive automated instrumentation for measuring ambient ozone. Air is sampled around a porous polypropylene tube filled with a solution of indigotrisulfonate. Light transmission through the tube is measured. Light transmission increases as O(3) diffuses through the membrane and bleaches the indigo. Evaporation of the solution, a function of the RH and the air temperature, can, however cause major errors. We solve this problem by adding an O(3)-inert dye that absorbs at a different wavelength. Here we provide a new algorithm for this correction and show that this very inexpensive instrument package (controlled by a BASIC Stamp Microcontroller with an on-board data logger, total parts cost US$ 300) provides data highly comparable to commercial ozone monitors over an extended period. The instrument displays an LOD of 1.2ppbv and a linear span up to 300ppbv for a sampling time of 1min. For a sampling time of 5min, the respective values are 0.24ppbv and 100ppbv O(3).