Reduction of axial kinetic energy induced perturbations on observed cyclotron frequency

J Am Soc Mass Spectrom. 2008 Apr;19(4):467-78. doi: 10.1016/j.jasms.2007.12.009. Epub 2008 Jan 6.

Abstract

With Fourier transform ion cyclotron resonance (FTICR) mass spectrometry one determines the mass-to-charge ratio of an ion by measuring its cyclotron frequency. However, the need to confine ions to the trapping region of the ion cyclotron resonance (ICR) cell with electric fields induces deviations from the unperturbed cyclotron frequency. Additional perturbations to the observed cyclotron frequency are often attributed to changes in space charge conditions. This study presents a detailed investigation of the observed ion cyclotron frequency as a function of ion z-axis kinetic energy. In a perfect three-dimensional quadrupolar field, cyclotron frequency is independent of position within the trap. However, in most ICR cell designs, this ideality is approximated only near the trap center and deviations arise from this ideal quadrupolar field as the ion moves both radially and axially from the center of the trap. To allow differentiation between deviations in observed cyclotron frequency caused from changes in space charge conditions or differences in oscillation amplitude, ions with identical molecular weights but different axial kinetic energy, and thus amplitude of z-axis motion, were simultaneously trapped within the ICR cell. This allows one to attribute deviations in observed cyclotron frequency to differences in the average force from the radial electric field experienced by ions of different axial amplitude. Experimentally derived magnetron frequency is compared with the magnetron frequency calculated using SIMION 7.0 for ions of different axial amplitude. Electron promoted ion coherence, or EPIC, is used to reduce the differences in radial electric fields at different axial positions. Thus with the application of EPIC, the differences in observed cyclotron frequencies are minimized for ions of different axial oscillation amplitudes.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Bradykinin / chemistry
  • Cyclotrons*
  • Insulin / chemistry
  • Melitten / chemistry
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Spectroscopy, Fourier Transform Infrared / instrumentation*
  • Spectroscopy, Fourier Transform Infrared / methods*
  • Tandem Mass Spectrometry
  • Thermodynamics

Substances

  • Insulin
  • Melitten
  • Bradykinin