A combined wireless neural stimulating and recording system for study of pain processing

J Neurosci Methods. 2008 May 15;170(1):25-34. doi: 10.1016/j.jneumeth.2007.12.014. Epub 2008 Jan 5.

Abstract

Clinical studies have shown that spinal or cortical neurostimulation could significantly improve pain relief. The currently available stimulators, however, are used only to generate specific electrical signals without the knowledge of physiologically responses caused from the stimulation. We thus propose a new system that can adaptively generate the optimized stimulating signals base on the correlated neuron activities. This new method could significantly improve the efficiency of neurostimulation for pain relief. We have developed an integrated wireless recording and stimulating system to transmit the neuronal signals and to activate the stimulator over the wireless link. A wearable prototype has been developed consisting of amplifiers, wireless modules and a microcontroller remotely controlled by a Labview program in a computer to generate desired stimulating pulses. The components were assembled on a board with a size of 2.5 cm x 5 cm to be carried by a rat. To validate our system, lumbar spinal cord dorsal horn neuron activities of anesthetized rats have been recorded in responses to various types of peripheral graded mechanical stimuli. The stimulation at the periaqueductal gray and anterior cingulate cortex with different combinations of electrical parameters showed a comparable inhibition of spinal cord dorsal horns activities in response to the mechanical stimuli. The Labview program was also used to monitor the neuronal activities and automatically activate the stimulator with designated pulses. Our wireless system has provided an opportunity for further study of pain processing with closed-loop stimulation paradigm in a potential new pain relief method.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Action Potentials / physiology
  • Animals
  • Cerebral Cortex / physiology
  • Electric Stimulation / instrumentation*
  • Electronics
  • Feedback / physiology
  • Male
  • Microelectrodes
  • Pain / physiopathology*
  • Pain Measurement / methods
  • Periaqueductal Gray / physiology
  • Physical Stimulation
  • Posterior Horn Cells / physiology
  • Rats
  • Rats, Sprague-Dawley
  • Software
  • Telemetry / instrumentation*