Performance of QPET, a high-resolution 3D PET imaging system for small volumes

IEEE Trans Med Imaging. 1994;13(1):176-85. doi: 10.1109/42.276156.

Abstract

QPET is a positron imaging system developed at Queen's University for high-resolution, 3D imaging of small volumes; it includes a pair of planar gamma-ray detectors 25.4 cm square, which rotate about a central axis with a quasi-cylindrical geometry. The authors describe the performance of this system. Basic characteristics of the detectors are evaluated: a spatial sampling of 1 mm, a quantum efficiency of 9.3% (for 511 keV gamma rays with normal incidence), and a time resolution of 88 ns. Models are developed to characterize the system deadtime and the sensitivity in terms of the noise-equivalent counting rate. With an 8 cm diameter spherical source, the noise-equivalent counting rate reaches a maximum at just over 3 kcps for an activity concentration of 2 muCi/cc; the random coincidence events and the deadtime losses both contribute significantly and the scatter contribution is small. Spatial resolution and uniformity over the field of view are evaluated by imaging short and long line sources; a spatial resolution of 2.7 mm in the transverse directions and 2.0 mm in the axial direction is achieved, with excellent uniformity throughout the field of view. The detector response is amplitude invariant across a 20 cm transverse diameter and a 9 cm axial length with the acceptance angle limited to +/-25 degrees in the axial direction. As an example of the imaging capabilities of QPET, the authors show 3D images of (18)F uptake in the bones of a rat, showing the excellent spatial resolution. This system is best suited to limited-volume applications where high counting rates are not necessary, but where high spatial resolution and uniform detector response are priorities.