Determination of picomole amounts of lipoxins C4, D4 and E4 by high-performance liquid chromatography with electrochemical detection

J Chromatogr. 1991 Oct 25;585(1):53-6. doi: 10.1016/0021-9673(91)85056-l.

Abstract

A method for the sensitive determination of the sulphopeptide lipoxins (LXs) C4, D4 and E4 by high-performance liquid chromatography with subsequent electrochemical detection is described. The best results were obtained when the analysis was carried out with the solvent system methanol-water-trifluoroacetic acid (66:34:0.008, v/v/v). The acquired half-wave potentials were different for all investigated compounds: +1.18 V for LXC4 +1.3 V for LXD4 and +1.25 V for LXE4. The detection limits of LXC4, LXD4 and LXE4, based on a signal-to-noise ratio of 3:1, were found to be 200-700 fmol. Although sulphopeptide lipoxins possess a high molar absorptivity, electrochemical detection still is three times more sensitive than ultraviolet detection.