Diagnostic accuracy of the Triage D-dimer test for exclusion of venous thromboembolism in outpatients

Thromb Res. 2008;121(6):735-41. doi: 10.1016/j.thromres.2007.07.012. Epub 2007 Sep 20.

Abstract

Background: We evaluated the diagnostic performance of the Triage D-dimer test, a new fast quantitative point-of-care whole blood D-dimer assay and compared it with the Vidas D-dimer assay.

Materials and methods: The study population comprised 319 outpatients for whom D-dimer testing was requested in order to rule out venous thromboembolism (VTE). Routine testing consisted of a plasma ELISA D-dimer analysis (Vidas). For all included patients, an additional EDTA whole blood D-dimer test (Triage) was performed. Patients were classified by reference imaging or by follow-up of the medical record. Accuracy indices, receiver operating characteristics and the kappa coefficient for agreement were calculated using the cutoff values recommended by the manufacturer.

Results: Prevalence of VTE was 14%. Sensitivity and specificity for VTE were 98% (95%CI: 88-100) and 34% (95%CI: 28-40) for Vidas and 91% (95%CI: 78-97) and 42% (95%CI: 36-48) for Triage, respectively. The differences in sensitivity and specificity between both D-dimer assays were statistically significant (McNemar, p<0.0001). ROC-curve analysis yielded an area under the curve of 0.83 (95%CI: 0.76-0.89) for the Vidas and 0.81 (95%CI: 0.74-0.88) for the Triage (p=0.396). The kappa coefficient for agreement between Vidas and Triage was 0.75 (95%CI: 0.68-0.79).

Conclusions: The Triage and Vidas D-dimer tests show comparable diagnostic accuracy. Vidas showed a significant higher sensitivity. Our findings strongly suggest lowering the cutoff for the Triage D-dimer test from 400 to 350 ng/mL. In this way specificity lowers from 42 to 38%, but, more importantly, sensitivity increases from 91 to 95%.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Diagnostic Techniques and Procedures / standards
  • Diagnostic Techniques and Procedures / statistics & numerical data
  • Female
  • Fibrin Fibrinogen Degradation Products / analysis*
  • Humans
  • Male
  • Middle Aged
  • Outpatients*
  • Point-of-Care Systems / standards
  • Predictive Value of Tests
  • Pulmonary Embolism / blood*
  • Pulmonary Embolism / diagnosis*
  • Regression Analysis
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Venous Thromboembolism / blood*
  • Venous Thromboembolism / diagnosis*

Substances

  • Fibrin Fibrinogen Degradation Products
  • fibrin fragment D