Studies on electrochemical behaviors of acyclovir and its voltammetric determination with nano-structured film electrode

Anal Chim Acta. 2006 Aug 18;576(1):17-22. doi: 10.1016/j.aca.2005.12.023. Epub 2006 Jan 31.

Abstract

A multi-wall carbon nanotubes (MWNTs)-dihexadecyl hydrogen phosphate (DHP) film-coated glassy carbon electrode (GCE) was fabricated, and the electrochemical behaviors of acyclovir on the MWNTs-DHP film-coated GCE were investigated by using cyclic voltammetry (CV), linear sweep voltammetry (LSV), electrochemical impedance spectroscopy (EIS) and chronocoulometry (CC). The oxidation peak current of acyclovir increased significantly and the peak potential shifted negatively at the MWNTs-DHP film-modified GCE, compared with that at a bare GCE. The results showed that this nano-structured film electrode exhibited excellent enhancement effects on the electrochemical oxidation of acyclovir. Consequently, a simple and sensitive electroanalytical method was developed for the determination of acyclovir. The oxidation peak current was proportional to the concentration of acyclovir from 8.0 x 10(-8) to 1.0 x 10(-5) mol/L. The detection limit was about 3.0 x 10(-8) mol/L for 60s accumulation at 0.00 V. The proposed method was demonstrated by using acyclovir tablets and the result was satisfying.