Elimination of caffeine interference in high-performance liquid chromatographic determination of cotinine in human plasma

J Chromatogr. 1991 Jul 17;568(1):267-9. doi: 10.1016/0378-4347(91)80363-h.

Abstract

A high-performance liquid chromatographic method with ultraviolet photometric detection for the determination of cotinine in human plasma was described. The use of a 30-cm reversed-phase column and of a mobile phase consisting of water-methanol-0.1 M sodium acetate-acetonitrile (72:21:5.6:1.4, v/v), pH 4.1, eliminated caffeine interference. A simplified solid-phase extraction procedure was also performed for plasma samples.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Caffeine / analysis
  • Caffeine / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Cotinine / analysis
  • Cotinine / blood*
  • Drug Interactions
  • Humans

Substances

  • Caffeine
  • Cotinine