SPY: an innovative intra-operative imaging system to evaluate graft patency during off-pump coronary artery bypass grafting

Interact Cardiovasc Thorac Surg. 2004 Sep;3(3):479-83. doi: 10.1016/j.icvts.2004.01.018.

Abstract

Off-pump coronary artery bypass grafting (CABG) has been rapidly increased, because of its less invasiveness with low complications. However, graft patency rate highly depends on the operators' capability due to technical difficulties. The SPY system, based on the fluorescence of indocyanine green, is an innovative device that permits validation of graft patency intra-operatively. Real time images of grafts are obtained with no need for catheterization, X-rays or iodine contrast medium. High-quality images could be obtained in all 290 grafts of 72 off-pump CABG cases (mean 4.0 grafts per patient). Four anastomoses (1.4%), including two proximal and two distal, were revised because of defects detected by SPY images. In one case, the SPY system revealed no blood flow in a radial sequential graft, although transit-time flow meter measurements showed a diastolic dominant pattern. SPY images provide critical information to surgeons to detect non-patent grafts, allowing them to be revised while the patient is still on the operating table. Using the SPY system, technical failures could be completely resolved during surgery. The use of the SPY system for intra-operative graft validation during off-pump CABG may become the gold standard for surgical management in the near future.