Numerical modelling errors in electrical impedance tomography

Physiol Meas. 2007 Jul;28(7):S45-55. doi: 10.1088/0967-3334/28/7/S04. Epub 2007 Jun 26.

Abstract

Electrical impedance tomography (EIT) is a non-invasive technique that aims to reconstruct images of internal impedance values of a volume of interest, based on measurements taken on the external boundary. Since most reconstruction algorithms rely on model-based approximations, it is important to ensure numerical accuracy for the model being used. This work demonstrates and highlights the importance of accurate modelling in terms of model discretization (meshing) and shows that although the predicted boundary data from a forward model may be within an accepted error, the calculated internal field, which is often used for image reconstruction, may contain errors, based on the mesh quality that will result in image artefacts.

Publication types

  • Validation Study

MeSH terms

  • Electric Impedance*
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Image Processing, Computer-Assisted / standards*
  • Models, Biological*
  • Reproducibility of Results
  • Tomography / methods
  • Tomography / standards*