High speed multi-frequency impedance analysis of single particles in a microfluidic cytometer using maximum length sequences

Lab Chip. 2007 Aug;7(8):1034-40. doi: 10.1039/b703546b. Epub 2007 Jun 8.

Abstract

A novel impedance spectroscopy technique has been developed for high speed single biological particle analysis. A microfluidic cytometer is used to measure the impedance of single micrometre sized latex particles at high speed across a range of frequencies. The setup uses a technique based on maximum length sequence (MLS) analysis, where the time-dependent response of the system is measured in the time domain and transformed into the impulse response using fast M-sequence transform (FMT). Finally fast Fourier transform (FFT) is applied to the impulse response to give the transfer-function of the system in the frequency domain. It is demonstrated that the MLS technique can give multi-frequency (broad-band) measurement in a short time period (ms). The impedance spectra of polystyrene micro-beads are measured at 512 evenly distributed frequencies over a range from 976.5625 Hz to 500 kHz. The spectral information for each bead is obtained in approximately 1 ms. Good agreement is shown between the MLS data and both circuit simulations and conventional AC single frequency measurements.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electric Impedance
  • Electrodes
  • Flow Cytometry / instrumentation
  • Flow Cytometry / methods*
  • Microfluidic Analytical Techniques / instrumentation
  • Microfluidic Analytical Techniques / methods*
  • Microspheres
  • Models, Theoretical*
  • Particle Size
  • Polystyrenes / chemistry

Substances

  • Polystyrenes