Optimization and validation of an automated voltammetric stripping technique for ultratrace metal analysis

Anal Chim Acta. 2007 Jul 2;594(2):192-8. doi: 10.1016/j.aca.2007.05.031. Epub 2007 May 26.

Abstract

A new automated batch method for the determination of ultratrace metals (nanogram per liter level) was developed and validated. Instrumental and chemical parameters affecting the performance of the method were carefully assessed and optimized. A wide range of voltammetric methods under different chemical conditions were tested. Cadmium, lead and copper were determined by anodic stripping voltammetry (ASV), while nickel, cobalt, rhodium and uranium by adsorptive cathodic stripping voltammetry (AdCSV). The figures of merit of all of these methods were determined: very good precision and accuracy were achieved, e.g. relative percentage standard deviation in the 4-13% for ASV and 2-5% for AdCSV. The stripping methods were applied to the determination of cadmium, lead, copper, nickel, cobalt, rhodium and uranium in lake water samples and the results were found to be comparable with ICP-MS data.