3D and 2D delayed-enhancement magnetic resonance imaging for detection of myocardial infarction: preclinical and clinical results

Acad Radiol. 2007 Jul;14(7):788-94. doi: 10.1016/j.acra.2007.03.006.

Abstract

Rationale and objectives: The purpose was to verify whether myocardial viability can be detected by a delayed enhancement magnetic resonance imaging (MRI) approach using a rapid three-dimensional inversion-recovery fast low-angle shot (3D IR-FLASH) sequence in a preclinical and clinical setting.

Materials and methods: Nonreperfused myocardial infarctions were induced in eight minipigs. Both the pigs and 15 patients with suspected myocardial infarction underwent MRI using a rapid 3D IR-FLASH sequence and a two-dimensional IR-FLASH sequence as the reference standard.

Results: In the pigs, a total of 52 segments with myocardial infarction were identified with both sequences and there was good agreement in transmurality of 99.5%. The infarction volume determined with the 3D IR-FLASH in the animal study (2.4 +/- 1.5 cm(3)) showed a good correlation with the histomorphometrically determined volume using triphenyltetrazolium chloride (2.3 +/- 1.2 cm(3), r = 0.98, P < .001) and the two-dimensional IR-FLASH sequence (2.3 +/- 1.4 cm(3), r = 0.99, P < .001). Eleven of 15 patients were found to have myocardial infarction in 37 myocardial segments with both sequences and there was a good agreement in transmurality of 98.8%. There was also a good correlation in the clinical study between the 3D and 2D sequences (6.9 +/- 6.7 cm(3) vs. 6.8 +/- 6.5 cm(3), r = 0.98, P < .001). In Bland-Altman analysis there was no significant under- or overestimation of the myocardial infarction volume using the 3D IR-FLASH sequence in comparison to the two-dimensional reference standard in both the preclinical and clinical study. The contrast-to-noise ratios were not significantly different between 3D and 2D sequences in the animal (34.7 +/- 1.5 vs. 33.8 +/- 2.6; P = .51) and clinical study (31.4 +/- 12.5 vs. 36.7 +/- 11.5; P = .31). The breathhold time for the 3D IR-FLASH sequence in the clinical study (20.4 +/- 2.2 s) was significantly shorter than that of the 2D IR-FLASH sequence (190.1 +/- 20.8 s, P < .001).

Conclusions: The rapid 3D IR-FLASH sequence detects myocardial infarction with high accuracy and allows a relevant reduction in acquisition time.

Publication types

  • Clinical Trial

MeSH terms

  • Aged
  • Animals
  • Chronic Disease
  • Contrast Media / administration & dosage
  • Female
  • Gadolinium DTPA
  • Heterocyclic Compounds
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Myocardial Infarction / diagnosis*
  • Organometallic Compounds
  • Predictive Value of Tests
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Swine
  • Swine, Miniature
  • Time Factors

Substances

  • Contrast Media
  • Heterocyclic Compounds
  • Organometallic Compounds
  • gadolinium 1,4,7,10-tetraazacyclododecane-N,N',N'',N'''-tetraacetate
  • Gadolinium DTPA